Free Call Australia : 1 800 834 991

Related Products

Omicron CAL 542 Charge Calibrator 

OMICRON CAL 542 Charge Calibrator 
  • Lightweight and easy to use and Amplitudes are easy to set up
  • Complies with all parts of IEC 60270 and Independent supply with built-in battery
  • Synchronization of the pulses to mains frequency with built-in light sensor

Product Details

OMICRON CAL 542 Charge Calibrator

Charge calibrator - Accessory for partial discharge measurement products

The IEC 60270 requires verifying the test setup prior to conducting measurements.
To achieve this, a defined charge pulse is fed into the measuring circuit with the help of the CAL 542. The partial discharge detection system is then calibrated to this value.
The CAL 542 is very easy to use. The charge value is shown on the display and can be adjusted using two buttons. Thanks to its battery power supply, the calibrator can be used at any time, anywhere.


Key features

• Lightweight and easy to use
• Amplitudes are easy to set up
• Complies with all parts of IEC 60270
• Synchronization of the pulses to mains frequency with built-in light sensor
• Independent supply with built-in battery


Applications

• For IEC 60270-compliant measurements
• Transformers, generators, cables, GIS, etc.

OMICRON CAL 543 Online Charge Calibrator Online charge calibrator - Accessory for partial discharge measurement products The CAL 543 is a charge injector, which remains connected to th...
OMICRON MONTRANO Continuous Online Monitoring System MONTRANO continuously monitors the dielectric condition of insulation in bushings and inside power transformers under load. The system...
PITE TECH KG2100B Transformer Dissolved Gas Monitor TransPRO® KG2100B is our classic model designed for continuous, long term and on-line monitoring of fault gases dissolved in oil. It is...
MEGGER PowerDB Test Data Management Software The Megger PowerDB is a powerful software package providing data management for each of your acceptance and maintenance testing jobs. Custo...
OMICRON PDL 650 Acoustic Partial Discharge Localization To take action on weak insulations quickly, the failure location must be known. Upcoming insulation defects already produce measura...